Guardian Co-Deposition Controller Operating Manual
3.3.5 Readings Panel—Readings Mode
Click View >> Readings to see the readings display. See
. The
readings display mode shows unfiltered channel (detector), sensor, and QCM 
readings.
Figure 3-10 Readings Panel Readings Display
3.3.5.1 Channel Readings
Each channel uses either two or three measurements to calculate the channel 
current signal. The measurements used are:
PMTCurOn: The first measurement cycle measures the detector current with 
sensor bias voltage on. This is a measure of the deposition flux, plus any 
background signal from ambient light or the sensor filament. This measurement 
also includes any broadband emission from other process gases.
PMTCurOff: The second measurement cycle measures the PMT current with 
sensor bias voltage off. With bias voltage off, no electron emission occurs. Hence, 
this is a measure of only background light.
PMTCurGas: The third measurement cycle is used only with INFICON's gas 
compensating (GC) sensor. The GC sensor includes an element that is not 
exposed to materials being deposited. It measures only the emission caused by 
gases, especially water vapor and nitrogen (plus any background light as explained 
above).
When the three measurements are completed, the channel readings are displayed 
as:
Flux - Gas: equals PMTCurOn minus PMTCurGas.