13 t, 14 v, 15 w – INFICON Guardian EIES Controller User Manual
Page 154
7 - 16
IP
N 07
4-
51
7-
P1
D
Guardian Co-Deposition Controller Operating Manual
7.3.4.13 T
Description: Reads voltage on one of the eight controller outputs. Value is returned
in DAC counts, with 0 = -10 V, 32767 = 0V, and 65535 = +10 V.
Parameters: Output#
Example: T1 (reads Output 1 voltage)
Returns: T
7.3.4.14 V
Description: Sets sensor emission current using a comma delimited string of four
current values: Sensor1 Current, Sensor1 Gas Current, Sensor2 Current, Sensor2
Gas Current. The emission current value is sent as mA x 100.
CAUTION
Refer to the label attached to your sensor or emitter
assembly package for correct emission current settings.
Do not operate 016-xxx series sensors with Yttria
filaments at Emission current greater than 2 mA. Instant
filament failure may occur! 016-xxx series sensors with
Thoria filaments may be operated at 4 mA emission.
Parameters: S1Current, S1GasCurrent, S2Current, S2GasCurrent
Example: V100,200,300,400 (sets Sensor1 to 1ma, Sensor1Gas to 2ma, etc.)
Returns: V
7.3.4.15 W
Description: Sets the output voltage of Guardian controller outputs 7 or 8. The
command is in DAC counts with 0 = -10V, 32767 = 0 V, and 65535 = +10 V.
See Note in
re "safe mode" timeout.
Parameters: Out# - 7,Counts
Example: W0,32767 (sets Output 7 to 0 V)
Returns: W