3 optical beam splitter, Section 4.3.3.3, Caution – INFICON Guardian EIES Controller User Manual
Page 103

4 - 11
IP
N 07
4-
51
7-
P1
D
Guardian Co-Deposition Controller Operating Manual
CAUTION
Be careful when bending fiber optic lines. Individual 
internal fibers are easily broken, which reduces the light 
signal at the output. Also, the detectors must be 
supported by an external means. Do not hang the 
detector from the Optical Beam Splitter fibers. Be sure 
power to the PMT Detector Module is turned off whenever 
the PMT is exposed to room light to avoid damage to the 
PMT.
4.3.3.3 Optical Beam Splitter
For 2-material systems, two detectors can be mounted at the two outlets of one 
beam-splitter. If there were no losses, half the original light signal would go to each 
detector. In reality, light losses are significant and much less than half actually goes 
to each detector. 
shows a simplified schematic of an optical beam
splitting system for two detectors.
Figure 4-11 Beam Splitter
Beam splitters from
are available with 1 in. (25 mm) threaded fittings to
mate with the Guardian detectors and flange couplings.
Multiple detectors and the associated optical components add size and weight to 
the detection system. Sufficient clearance and support must be provided for all 
these components. They cannot be supported by the vacuum feedthrough as in the 
one-detector case. When installing the detector system, be sure to allow clearance 
for maintenance of the sensor/feedthrough assembly.
For 3-material systems the signals reaching all three detectors are not the same 
magnitude. Because of the additional beam split, detectors #2 & #3 each get much 
less light than detector #1. The Guardian controller and software provide for gain 
beam
splitter
optical
detect
PM
extension
tube
optical
detect
PM