N 07
4-
51
7-
P1
D
Guardian Co-Deposition Controller Operating Manual
WARNING
Maintain adequate insulation and physical separation of 
sensor, detector, and I/O wiring from hazardous voltages.
WARNING
Relay contacts are rated for 30 V (dc) maximum duty. Any 
customer supplied relay contact circuit must be fused at 
not more than 5A if the circuit is capable of supplying 
more than 5A, including the available short circuit 
current.
2.2.5 QCM Option Installation
If an SQM-242 or Q-pod will be used for calibration or control, follow its Operating 
Manual to install the QCM software and sensor(s).
For accurate rate calibration, locate the QCM sensor(s) in the vacuum chamber so 
that the QCM sensor(s) receives a representative sample of the materials being 
deposited.
2.3 EIES Software
Leave the Guardian Controller power OFF while you become familiar with the EIES 
program. This section will introduce the concepts necessary to operate the EIES-IV 
Guardian system.
2.3.1 Software Installation
Insert the INFICON Program Disk into your CD-ROM drive. The Products Menu 
should appear after a few seconds, see 
. If not, open
Windows Explorer and run the UtilityDisk.exe program on the CDROM.
On the Products Menu, click EIES-IV Guardian. When EIES software installation 
finishes, you may be prompted to restart your computer.