3 zero calibration, 4 gain calibration - test run, Cal. gain measured thickness eies total – INFICON Guardian EIES Controller User Manual
Page 133

6 - 9
IP
N 07
4-
51
7-
P1
D
Guardian Co-Deposition Controller Operating Manual
On the Edit Layer dialog box, adjust the PMT voltage until the % Current reading 
on the Edit System dialog box shows 50% to 70%. Typical PMT voltage values are 
500 V to 700 V. The 50% to 70% value may seem low, but some headroom may 
be necessary when other materials are being codeposited.
6.4.3 Zero Calibration
The EIES detector may read a residual gas signal, even when no deposition is 
occurring. Allow a one hour warm-up for the PMT Detector module.
To remove this error term, in the EIES software create a layer that uses the desired 
material (film). Show the channel's Edit Layer dialog box, Film Cal. tab. Verify that 
the film's User Multiplier setting is 1 and User Offset is 0. Also verify that the Cal 
Gain setting is 1 and the Cal Zero setting is 0. This assures that the detector 
reading displayed is detector current in nA.
With the EIES sensor enabled, but with no deposition, observe the channel 
readings. Enter the reading into the Cal. Zero field of the Edit Layer dialog box and 
press Enter. The channel reading will now show zero.
6.4.4 Gain Calibration - Test Run
Gain calibration matches the EIES detector's relative rate/thickness measurement 
to a known rate/thickness measurement. This can be done by completing a trial 
deposition on a substrate whose thickness can be measured by some other 
means, such as a QCM, interferometer, profilometer, or weight. Allow a one hour 
warm-up for optimum stability.
Record the EIES thickness reading at the end of the deposition and correct the 
value by entering a Cal Gain:
[13]
CAUTION
This method of EIES gain calibration is both 
time-consuming and prone to error.
Cal. Gain
Measured Thickness
EIES Total
-------------------------------------------------
=