Equipment options
Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

78

Abb. 12-6: Hydrophobic filter and holder assembly (exploded view)

Fig. 12-6: Hydrophobic filter and holder assembly (exploded view)

5. Make sure the O-rings (between the cap & filter holder and between the filter holder & oxygen sensor

holder) are undamaged and securely in place.

6. Gently press the filter cap back in.

7. Press fit the filter assembly back onto the oxygen sensor holder.

12.4.9

Troubleshooting the O

2

sensor

If the oxygen sensor fails suddenly, it is very likely that the sensor inlet membrane has become blocked by
condensation. This can be seen on the

DATALOGGER screen as a sudden drop from the programmed value

to nearly zero.

To dry the membrane:

1. Remove the hydrophobic filter holder assembly (see Fig. 12-4 on p. 76) by unscrewing it

(counter-clockwise).

2. Program the incubator for a temperature of at least 37 °C (or higher if you normally operate the

incubator at a higher temperature).

3. Close the door and allow the temperature to recover.

1

Oxygen sensor holder

2

Hydrophobic filter disc

3

Hydrophobic filter holder

4

Hydrophobic filter cap

5

Oxygen sensor