5 datalogger screen, 1 alarm events, Datalogger screen 6.5.1 – Eppendorf Galaxy CO2 Sensor User Manual
Page 36: Alarm events, Or detailed information (see, Datalogger screen on p. 36)

Operating controls and function for Galaxy 170 R Incubator
Galaxy
®
170 R/170 S CO
2
Incubators
English (EN)
36
disinfected prior to resumption of activity. For a full explanation of this feature, (see
disinfection option with oxygen control on p. 71).
6.5
DATALOGGER
screen
The
DATALOGGER
screen displays the following information:
Abb. 6-4:
DATALOGGER screen
Fig. 6-4:
DATALOGGER
screen
6.5.1
ALARM EVENTS
The following alarm events are recorded in the order in which they occurred, with the most recent event
displayed at the top:
• Power ON/OFF
• Chamber Temperature High/Low (programmed value)
• CO
2
Level High/Low (programmed value)
• CO
2
Supply Failure
• All System Alarms
• CO
2
Auto-Zero (A/Z) Adjustments
• CO
2
Auto Gain (A/G) Adjustments (
reserved for use by authorized service technicians only)
• Oxygen and Relative Humidity (R/H) Alarms (where these options are installed)
The capacity is 99 events, after which the earliest event is overwritten and a later event is added.
The date and the time are also recorded for each event, (see Fig. 6-5 on p. 37):