6.5.5
DIAGNOSTIC DOOR GRAPH
This graph records the door’s inner surface temperature over time to assist troubleshooting.
6.5.6
DIAGNOSTIC DOOR ELEMENT GRAPH
This graph records door element temperature over time to assist troubleshooting.
6.5.7
RESTART GRAPHIC RECORD
This feature removes the current graph and begins a new one. The data cannot be recovered once it is
deleted.
6.6
CHAMBER ALARMS
menu screen
The
CHAMBER ALARMS
programming screen (see Fig. 6-8 on p. 39) allows the various alarm options to be
selected and modified. Press the
or
direction key to move around the options and the
or
direction key to adjust values. The temperature and CO
2
High and Low Alarm setpoints automatically adjust
to within ± 0.5 of the temperature and CO
2
setpoints. The alarm setpoints can also be manually adjusted.
Abb. 6-8:
CHAMBER ALARMS screen
Fig. 6-8:
CHAMBER ALARMS
screen
To arm the chamber alarms after a selectable delay:
1. Choose the option
ARM ALARMS WHEN AT SETPOINT
.
2. Select
NO
for both
TEMP
and
CO
2
3. Choose the option
DELAY IN ARMING AFTER DOOR OPEN
and select the desired delay (15 minutes in
the sample screen (Fig. 6-8 on p. 39)) to allow for temperature and CO
2
recovery after the door has been
opened.
Alternatively, the alarm system can be set to re-arm only after the original temperature and CO
2
setpoints
have been achieved:
1. Choose the option
ARM ALARMS WHEN AT SETPOINT
.