Operating controls and function for Galaxy 170 R Incubator

Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

39

6.5.5

DIAGNOSTIC DOOR GRAPH

This graph records the door’s inner surface temperature over time to assist troubleshooting.

6.5.6

DIAGNOSTIC DOOR ELEMENT GRAPH

This graph records door element temperature over time to assist troubleshooting.

6.5.7

RESTART GRAPHIC RECORD

This feature removes the current graph and begins a new one. The data cannot be recovered once it is
deleted.

6.6

CHAMBER ALARMS

menu screen

The

CHAMBER ALARMS

programming screen (see Fig. 6-8 on p. 39) allows the various alarm options to be

selected and modified. Press the

or

direction key to move around the options and the

or

direction key to adjust values. The temperature and CO

2

High and Low Alarm setpoints automatically adjust

to within ± 0.5 of the temperature and CO

2

setpoints. The alarm setpoints can also be manually adjusted.

Abb. 6-8:

CHAMBER ALARMS screen

Fig. 6-8:

CHAMBER ALARMS

screen

To arm the chamber alarms after a selectable delay:
1. Choose the option

ARM ALARMS WHEN AT SETPOINT

.

2. Select

NO

for both

TEMP

and

CO

2

(see Fig. 6-8 on p. 39).

3. Choose the option

DELAY IN ARMING AFTER DOOR OPEN

and select the desired delay (15 minutes in

the sample screen (Fig. 6-8 on p. 39)) to allow for temperature and CO

2

recovery after the door has been

opened.

Alternatively, the alarm system can be set to re-arm only after the original temperature and CO

2

setpoints

have been achieved:
1. Choose the option

ARM ALARMS WHEN AT SETPOINT

.