Equipment options

Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

67

Abb. 12-1: BMS relay contact alarm socket

Fig. 12-1: BMS relay contact alarm socket

*Cable length should not exceed 3 m (9.8 ft) to comply with EMC requirements.

The default setting for the alarm system is ON. To deactivate the relay using the incubator keypad:

1. Press USER.

2. Select BMS ALARM RELAY.

3. Select MAKE ALARM RELAY ACTIVE YES/NO.

4. Toggle to NO and then press ENTER.

The default setting for the power failure warning is ON. To make the alarm system ignore any power
outage:

1. Press USER.

2. Select BMS ALARM RELAY.

3. Select MAKE ALARM RELAY ACTIVE AT POWER SWITCH OFF/FAILURE YES/NO.

4. Toggle to NO and then press ENTER.

Pin

Designation

1

24 V DC unregulated (170 R Models)
12 V DC unregulated (170 S Models)

To power external equipment such as a remote buzzer
or light (100 mA maximum current available).*

2

0 V

3

5 V DC

Via 10 K Ω pull-up resistor, for a logic signal to
directly control an auxiliary control system.*

4

Normally closed

To access the relay contacts. Contact limits are 3
Amps @ 24 V DC and 3 Amps @ 34 V AC.

5

Common

6

Normally open