beautypg.com

7 removing and replacing o2 sensor, Removing and replacing o, Sensor – Eppendorf Galaxy CO2 Sensor User Manual

Page 76: Ovided (see, Sensor on p. 76), Eplacing the sensor (see

background image

Equipment options
Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

76

OXYGEN CONTROL HAS BEEN DISABLED AS A RESULT BUT THE INCUBATOR IS OTHERWISE FULLY
OPERATIONAL.

PRESS PREV TO VIEW PREVIOUS SCREEN. PRESS EXIT TO EXIT.

When you press the

EXIT function key, you will return to the USER screen and normal operation.

12.4.7

Removing and replacing O

2

sensor

Prerequisites

• Oxygen sensor removal tool (see Fig. 12-5 on p. 77)

1. Pull the rear access cover (see Fig. 12-4 on p. 76) off the rear outside wall of the incubator to gain

access to the oxygen sensor.

Abb. 12-4: Oxygen sensor rear access cover

Fig. 12-4: Oxygen sensor rear access cover

2. Reach inside and disconnect the sensor by unplugging the connector: be sure to grasp the white

connector body.

3. Using the sensor removal tool (see Fig. 12-5 on p. 77), unscrew the oxygen sensor by turning it

counter-clockwise. Because the oxygen sensor contains lead, be sure to dispose of it according to local
regulations.

NOTICE! Risk of material damage

Grasp white connector body when disconnection wire leads.

Do not pull on wire leads.