Operating controls and function for Galaxy 170 R Incubator

Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

33

6.3

Referencing the CO

2

sensor with Auto-Zero

Prior to using the incubator, you should manually perform a CO

2

Auto-Zero (see

PROGRAMMABLE CO

2

AUTOZERO on p. 34):

1. Perform a CO

2

Auto-Zero by pressing the USER function key (see

PROGRAMMABLE CO

2

AUTOZERO on

p. 34), selecting the PROGRAMMABLE CO

2

AUTOZERO, and pressing the START key.

2. The incubator will display a countdown as the Auto-Zero is running.

3. When the countdown is complete, the incubator is ready to use.

6.4

USER settings

In the

USER screen, you can adjust the features called out on the screen.

Abb. 6-2: USER SETTINGS screen

Fig. 6-2:

USER SETTINGS screen

This section explains each of the

USER screen features. There are other USER options that may be

displayed on screen if they are installed on your incubator. For a list of available options (see

Available

options on p. 64).

If the access code has been misplaced, you will be unable to make changes to your incubator’s
settings. Contact customer service or your service representative for instructions on how to
regain access to your incubator.

1

Use the

and

direction keys to move the

cursor

2

Use the ENTER function key to select an option

2