2 service, 1 fuse replacement, 3 cleaning – Eppendorf Galaxy CO2 Sensor User Manual
Page 53: 4 disinfection/decontamination, Service 8.2.1, Fuse replacement, Cleaning, Disinfection/decontamination

Maintenance
Galaxy
®
170 R/170 S CO
2
Incubators
English (EN)
53
8.2
Service
8.2.1
Fuse replacement
To replace a fuse, you will need a flat-bladed (5 mm maximum) screwdriver.
1. Using the flat-bladed screwdriver, rotate the fuse holder (see Fig. 4-7 on p. 27) counter-clockwise until
the fuse holder springs out.
2. Remove the spent fuse.
3. Install a new fuse of the same type.
4. Orient the fuse holder back in the side panel and, using the screwdriver, turn clockwise to secure it in
place.
8.3
Cleaning
1. Routinely clean the exterior of the incubator by wiping it over with a soft cloth, moistened with soapy
water.
2. Rinse the soap from the cloth in clean water, and wipe the exterior surfaces again.
8.4
Disinfection/Decontamination
DANGER! Electric shock
Switch the device off and pull the power plug out of the socket before beginning work.
DANGER! Electric shock
Switch the device off and pull the power plug out of the socket before beginning work.
NOTICE! Damage due to incorrect cleaning agent or sharp objects
Unsuitable cleaning agents can damage the display, surfaces and printing.
Never use corrosive cleaning agents, strong solvents or abrasive polishes.
Do not use acetone to clean the device.
Do not use sharp objects to clean the device.
DANGER! Danger from exposure to decontamination agents.
Wear appropriate laboratory clothing, protective gloves and safety glasses.
Wear breathing protection if you work with particulate matter.