Installation

4.1

Utilities requirements

The following utilities requirements are needed for operation:

4.2

Selecting the location

Select a level surface capable of withstanding the operating weight of the incubator. Actual operating
weight will be dependent on both the options installed, and the material stored in the incubator.

The incubator is designed to operate at a chamber temperature of 4.0 °C above ambient, and at an absolute
minimum ambient temperature of 15 °C if the incubator is being operated at 37 °C.

Maximum allowable

ambient temperature is 28 °C.

Avoid placing the incubator in areas that may affect performance, such as those listed below.

DO NOT place the incubator:

• Directly under, beside or within the air flow of heating or air-conditioning ducts, or other drafts;
• Directly beside heat-generating equipment such as a heater, an autoclave or an oven;
• Near the exhaust of heat- or cold-generating equipment;
• Near a window exposed to direct sunlight;
• Directly on top of any heat-generating apparatus;
• Without minimum ventilation clearance of 10 mm (0.5 in) all around (50 mm (2 in) in back if you have

the cooling system option).

Utility

Requirement

Electricity

120

V, 50/60

Hz earthed/grounded mains/electrical supply with minimum

capacity of 10

amps

230

V, 50/60

Hz earthed/grounded mains/electrical supply with minimum

capacity of 8

amps

CO

2

gas

Cylinder with 100

%

CO

2

vapor withdrawal, together with a two-stage

regulator for pressure control to 18.85

PSI

(1.3

bar)

CO

2

gas pressure must not exceed 5

PSI

(0.35

bar). It is recommended that a in-line pressure

regulator be used for each gas type introduced into the incubator to precisely control gas
pressure (see

Accessories on p. 63).

Position incubator to allow clearance for opening door and access to the CO

2

sample port

located on the left side of the incubator.