Operating controls and function for Galaxy 170 R Incubator
Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

32

6.2

Programming

6.2.1

Temperature and CO

2

level

Perform the following steps to set the desired operating temperature and CO

2

level. For more information

on this feature (see

Infrared sensor on p. 13).

1. Press the

PROG function key.

2. In the

PROG screen that appears, press the desired function key, TEMP or CO

2

, then use the

and

direction keys to adjust the value.

3. When the desired setpoint is displayed, press the

ENTER function key.

4. After making adjustments (if any were made), allow the incubator to stabilize at the setpoints before

continuing.

6.2.2

User access code

Programmable user access code allows you to restrict access to the

PROG, USER, and ALARM screens

(where settings can be changed) to authorized persons only.

To set the User Access Code (if required):

1. In the

PROG screen (accessed by pressing the PROG function key).

The user access code will be displayed as a series of four asterisks.

2. Use the left and right direction keys to move to each code position, and the up and down direction keys

to select a number from 0 to 9.

3. Once the number is selected, press the

ENTER function key to save the code.

4. After returning to the main screen, programming access will require the code to make any further

programming changes.

6.2.3

Removing user access code

1. In the

PROG screen, enter the current access code.

2. Now program

0000 as the new access code.

3. Press the

ENTER function key to save the change.

The code is now cancelled and programming is no longer restricted.

If the incubator is supplied with the option of oxygen control, the setpoint for the oxygen level
can be selected and changed like the temperature and CO

2

setpoints.

If the chamber temperature goes above the temperature setpoint by 1 °C, the
over-temperature system will activate.

Take care to note your password somewhere. If a password is forgotten, you must contact a
customer service representative to recover or delete the forgotten password.