Equipment options

Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

73

4. After the incubator has been humidified and left overnight to stabilize, select the

USER menu; then,

using the

or

direction key, select

OXYGEN SENSOR-REF TO ATMOSPHERE (see Fig. 12-2 on

p. 73) and follow the onscreen instructions to automatically calibrate the oxygen sensor to atmospheric
oxygen levels. The oxygen reading is automatically adjusted to 19.7 %, which is the true reading taking
into account the relative humidity level.

Abb. 12-2: Selecting OXYGEN SENSOR-REF TO ATMOSPHERE

Fig. 12-2: Selecting OXYGEN SENSOR-REF TO ATMOSPHERE

5. Enable the oxygen control: (a) press the

USER function key, (b) using the

or

direction key, select

MANUAL DISABLE (see Fig. 12-3 on p. 73), (c) press the ENTER function key, (d) select ENABLE for

Oxygen Control using the

or

direction key, then (e) press the

ENTER function key.

Abb. 12-3: USER SETTINGS Screen

Fig. 12-3: USER SETTINGS Screen

6. Navigate to the

PROG screen to set the required Oxygen level.

7. The Alarm levels are automatically set to ± 0.5 % above or below the programmed value, but you can

modify them in the ALARM screen. Re-arming the alarm can be delayed until the programmed value is
achieved: select the relevant option in the ALARM screen.

If the programmed O

2

level is close to the ambient oxygen, it may be necessary to reduce the

cylinder pressure below 1 bar to stop the oxygen level from undershooting the programmed
value.
Both inner and outer doors must be open.