Safety

Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

19

NOTICE! Risk of material damage

CO

2

gas pressure must not exceed 5 PSI (0.35 bar).

NOTICE! Risk of material damage
Working with electrical power inside a humid environment (where the incubator is
humidified) can cause damage. The following precautions should be observed:

The instrument or equipment, and its external connections, to be used inside the chamber
should be specified as suitable for use in a humid environment, and at 37 °C (see also
“Using Powered Equipment within the Chamber”). If in doubt, consult with the
manufacturer of the equipment.

Always ensure the connections are properly and securely made.

Be sure to switch OFF the green illuminated switch on the front left of the incubator before
connecting or disconnecting equipment inside the chamber, if equipped with optional IP66
socket.

The Sealing Cap must always be in place when the socket is not in use.

Both the incubator and the IP66 enclosure must be plugged into an electrical supply
protected by an RCD device. Any device chosen must be a self-resetting type which will
automatically reconnect power to the incubator as soon as power is restored following a
power failure.