Equipment options

Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

75

12.4.5

Replace sensor soon

If the referencing procedure was successful but the sensor is nearing the end of its working life, the
following message will appear in the display:

O

2

REFERENCE OK BUT SENSOR REQUIRES REPLACEMENT SHORTLY

PRESS ENTER TO PROCEED

When you press the

ENTER function key, the message will change to this:

O

2

SENSOR

THE RESULT OF THE O

2

REFERENCE PROCESS SHOWS THAT THE SIGNAL FROM THE O

2

SENSOR HAS

REDUCED INDICATING IT IS APPROACHING THE END OF ITS LIFE

.

REPEAT THE REFERENCE PROCEDURE TO CONFIRM THIS RESULT

.

PRESS ENTER TO PROCEED

.

Press the

ENTER function key.

12.4.6

Replace sensor now

If the referencing procedure failed, Oxygen Control will be disabled. The incubator will appear to be
functioning as normal until a new sensor is installed and referenced to atmospheric level. The following
message will appear in the display:

O

2

REFERENCE FAILED

PRESS ENTER TO PROCEED

When you press the

ENTER function key, the message will change to this:

O

2

SENSOR

THE RESULT OF THE O

2

REFERENCE PROCESS SHOWS THAT THE SIGNAL FROM THE O

2

SENSOR HAS

REDUCED BELOW AN ACCEPTABLE LEVEL AND HAS REACHED THE END OF ITS LIFE.

REPEAT THE REFERENCE PROCEDURE TO CONFIRM THIS RESULT.

PRESS NEXT TO PROCEED.

When you press the

NEXT function key, the message will change to this:

O

2

SENSOR