5 replace sensor soon, 6 replace sensor now, Replace sensor soon – Eppendorf Galaxy CO2 Sensor User Manual
Page 75: Replace sensor now, Esented in detail in (see, Replace sensor soon on p. 75) and (see, Replace sensor now on p. 75)

Equipment options
Galaxy
®
170 R/170 S CO
2
Incubators
English (EN)
75
12.4.5
Replace sensor soon
If the referencing procedure was successful but the sensor is nearing the end of its working life, the
following message will appear in the display:
O
2
REFERENCE OK BUT SENSOR REQUIRES REPLACEMENT SHORTLY
PRESS ENTER TO PROCEED
When you press the
ENTER function key, the message will change to this:
O
2
SENSOR
THE RESULT OF THE O
2
REFERENCE PROCESS SHOWS THAT THE SIGNAL FROM THE O
2
SENSOR HAS
REDUCED INDICATING IT IS APPROACHING THE END OF ITS LIFE
.
REPEAT THE REFERENCE PROCEDURE TO CONFIRM THIS RESULT
.
PRESS ENTER TO PROCEED
.
Press the
ENTER function key.
12.4.6
Replace sensor now
If the referencing procedure failed, Oxygen Control will be disabled. The incubator will appear to be
functioning as normal until a new sensor is installed and referenced to atmospheric level. The following
message will appear in the display:
O
2
REFERENCE FAILED
PRESS ENTER TO PROCEED
When you press the
ENTER function key, the message will change to this:
O
2
SENSOR
THE RESULT OF THE O
2
REFERENCE PROCESS SHOWS THAT THE SIGNAL FROM THE O
2
SENSOR HAS
REDUCED BELOW AN ACCEPTABLE LEVEL AND HAS REACHED THE END OF ITS LIFE.
REPEAT THE REFERENCE PROCEDURE TO CONFIRM THIS RESULT.
PRESS NEXT TO PROCEED.
When you press the
NEXT function key, the message will change to this:
O
2
SENSOR