2 power supply, 1 mains/electrical supply, 3 fuses – Eppendorf Galaxy CO2 Sensor User Manual
Page 60: 4 ambient conditions, 1 temperature management, Power supply 10.2.1, Mains/electrical supply, Fuses, Ambient conditions 10.4.1, Temperature management
Technical data
Galaxy
®
170 R/170 S CO
2
Incubators
English (EN)
60
10.2
Power supply
10.2.1
Mains/electrical supply
10.3
Fuses
10.4
Ambient conditions
10.4.1
Temperature management
• Digital programming via microprocessor control in 0.1 °C increments. Measurement of chamber and
door temperature via 6 RT (Resistance Temperature curve) matched thermistors (sensitivity 0.01 °C)
• Adjustable independent control of door heater
• “Out of Limits” temperature protection system independent of microprocessor control.
Mains/power connection
120 V, 50 Hz
120 V, 60 Hz
230 V, 50 Hz
230 V, 60 Hz
Power consumption for 120 V
1300 W
Power consumption for 230 V
1600 W
Energy to maintain 37 °C
< 0.1 kWh
fuse for 120 V
10 A
fuse for 230 V
8 A
Ambient conditions
15 °C - 28 °C
Range
4 °C above ambient temperature to 50 °C
Control
± 0.1 °C
Stability
± 0.1 °C at 37 °C
Uniformity
± 0.3 °C at ambient 20°C - 25 °C
If ambient temperature is close to the programmed value, control settings may need adjusting.
Please consult Eppendorf Service for instructions.