Equipment options

Galaxy

®

170 R/170 S CO

2

Incubators

English (EN)

69

was completed successfully,

DISINFECTION COMPLETED OK

is shown.

was cancelled by the user,

DISINFECTION WAS ABORTED

is shown.

failed for any reason,

DISINFECTION FAILED [CODE: XX]

is shown.

The following tables lists the disinfection failure codes and descriptions, (see Tab. on p. 69) and (see
Tab. on p. 69). If this happens, note the failur
e code and contact your service representative for advice.
Disinfection failure codes and descriptions

Disinfection failure code explanations

Failure Code

Failure Code Description (see Tab. on p. 69)

01

Z

02

W

03

W, Z

04

X

05

X, Z

06

W, X

07

W, X, Z

08

Y

09

Y, Z

0A

W, Y

0B

W, Y, Z

0C

X, Y

0D

X, Y, Z

0E

W, X, Y

0F

W, X, Y, Z

Failure Code Description

Explanation

W

Temperature drop during warm-up period: indicates the temperature fell
more than 2

°C during the heating phase over a 60-second period.

X

Temperature drop during 4-hour period: indicates the temperature fell
below 118.0

°C during the disinfection phase.

Y

Temperature increase during cool-down phase: indicates the temperature
rose by more than 2

°C during the cooling phase over a 60-second period.

Z

Cancel key pressed.

• If the incubator power is cycled OFF then ON during a disinfection cycle due to a power

outage, the incubator will power up as normal. This condition will be indicated by the
absence of a completed disinfection status message (

DISINFECTION COMPLETED OK

).

• If the chamber temperature is above the setpoint or the element temperature is greater

than a factory-preset control point, cool down will be entered until these conditions are
satisfied.

• It is recommended that the Auto-Zero function be run following each disinfection cycle.