2 setting the co2 high and low alarms, 3 door open alarm, 4 alarm duration – Eppendorf Galaxy CO2 Sensor User Manual
Page 45: Setting the co, High and low alarms, Door open alarm, Alarm duration

Operating controls and function for Galaxy 170 S Incubator
Galaxy
®
170 R/170 S CO
2
Incubators
English (EN)
45
6. Press the
Enter key to accept the setting.
7. Press the
Programming ( * ) key twice to exit the menu.
7.3.2
Setting the CO
2
high and low alarms
1. Press the
Enter and
keys simultaneously to enter the alarm menu.
The display will show:
°C.AL
.
2. Press the
key until the display shows:
CO2AL
.
3. Press the
Enter key to display
HI.5.5
. The factory setting is the setpoint value (5.0 %) + 0.5 %.
4. If you wish to adjust the High CO
2
Alarm, use the
or
key. The minimum setting is 0.5 % from
setpoint.
5. Press the
Enter key to save the setting.
The low CO
2
Alarm is displayed,
LO 4.5
. The factory setting is the setpoint value (5.0 %) - 0.5 %.
6. Use the the
or
key to adjust the value.
7. Press the
Enter key to accept the setting.
8. Press the
Programming ( * ) key twice to exit the menu.
7.3.3
Door open alarm
When the door is opened, an alarm will sound after a preset time delay. To adjust the time delay:
1. Press the
Enter and
keys simultaneously to enter Alarm Program Mode.
The display will show:
°C.AL
.
2. Press the
key until the display shows
DOO.RAL
(reading across both displays).
3. Press the enter key and the
and
keys to adjust the time (as you scroll through the available choices,
you will see
15, 30, 45, 60, 75, 90
seconds, then
OFF
).
4. Press the
Enter key to save the desired value.
5. Press the
Programming ( * ) key twice to exit the menu.
7.3.4
Alarm duration
Perform the following steps to adjust the alarm duration:
1. Press the
Enter and
keys simultaneously to enter Alarm Program Mode.
The display will show:
°C.AL
.
2. Press the
key until the display shows
PER.IOD
(reading across both displays).
3. Press the
Enter key and the
and
keys to adjust the Alarm duration.
If the CO
2
setpoint is programmed at 0.0 % and high and low alarms are accessed, the high
will display “0.5” and the low will display “Off”.