Eppendorf Innova 43R Incubator Shakers User Manual
Page 79
79
New Brunswick
Operating manual
Code Meaning
Format
PR Read
Profile
step in
memory
PR<Space><Arg1><Space><Arg2><CR>
where Arg1= Profile number (range 1 to 4) and
Arg2=Step number (range 1 to 15)
Return data format:
<Arg1><Space><Arg2><Space><Arg3><Space>...<CR><LF>
where Arg1=Profile number (range 1 to 4),
Arg2=Step number,
Arg3=Temperature setpoint,
Arg4=Agitation setpoint,
Arg5=(Reserved, reads 0.0),
Arg6=Step hours,
Arg7=Step minutes,
Arg8=UV lamp, and
Arg9=Grow lamp
PS Profile
Stop/
Start
(stops or
starts a
program
profile)
PS<CR>
With no other parameters, any profile currently running will be
stopped.
PS<Space><Arg1><Space><Arg2><CR>
where Arg1=Profile number (range 1 to 4), and
Arg2=Step number (range 1 to 15).
NOTE: If Arg2 is not included, the profile will start in Step 1.
PW Write
Profile
step to
memory
PW<Space><Arg1><Space><Arg2><Space><Arg3><Space>
...<CR>
where Arg1=Profile number (range 1 to 4),
Arg2=Step number (range 1 to 15),
Arg3=Temperature setpoint,
Arg4=Agitation setpoint,
Arg5=(Reserved); Enter 0.0,
Arg6=Step hours (range 0 to 99),
Arg7=Step minutes (range 0 to 59),
Arg8=UV lamp (0 = off, 1 = on), and
Arg9=Grow lamp (0 = off, 1 = on)
For example, the following
PW
command:
PW<Space>25.0<Space>150<Space>0.0<Space>2<Space>30<Space>0
<Space>1<CR>
will set Program 1, Step 1 to the following conditions:
Temperature setpoint 25.0º
Step Time 2 hours 30 minutes
Agitation setpoint 150 rpm
UV Lamp OFF
GRO Lamp ON