Eppendorf Innova 43R Incubator Shakers User Manual
Page 78

78
Innova 43 & 43R Incubator Shakers
M1320-0050
Operating manual
12.4 Set
Commands
Code
Meaning
Format
CL Command
Grow
Lamp
CL<Space><Arg1><CR>
where Arg1= Set Grow Lamp state (0 = off, 1 = on)
CS Command
Speed CS<Space><Arg1><CR>
where Arg1= Agitation setpoint (####)
CT Command
Temperature
CT<Space><Arg1><CR>
where Arg1= Temperature setpoint (###.#)
CU Command
UV
Lamp
CU<Space><Arg1><CR>
where Arg1= Set UV Lamp state (0 = off, 1 = on)
12.5
Profile Control Commands
Code
Meaning
Format
PC Clear
Profile
(clears profile
step)
PC<Space><Arg1><Space><Arg2><CR>
where Arg1= Profile number (range 1 to 4) and
Arg2=Step number (range 1 to 15)
NOTE: If Arg2 is not included, all 15 steps of the
selected profile will be cleared.
PM Profile
Mode
(returns status of
current profile
run/hold mode)
PM<CR>
There are no parameters to this command.
Return data format:
<Arg1><Space><Arg2><Space><Arg3><CR>
where Arg1=Run/hold status (0 = end, 1 = run),
Arg2=Profile number and
Arg3=Step number
NOTE: If a profile has not been selected to run, a 0
will be returned for Profile number and Step number.
...continued...