beautypg.com

Eppendorf Innova 43R Incubator Shakers User Manual

Page 78

background image

78

Innova 43 & 43R Incubator Shakers

M1320-0050

Operating manual

12.4 Set

Commands

Code

Meaning

Format

CL Command

Grow

Lamp

CL<Space><Arg1><CR>
where Arg1= Set Grow Lamp state (0 = off, 1 = on)

CS Command

Speed CS<Space><Arg1><CR>

where Arg1= Agitation setpoint (####)

CT Command

Temperature

CT<Space><Arg1><CR>
where Arg1= Temperature setpoint (###.#)

CU Command

UV

Lamp

CU<Space><Arg1><CR>
where Arg1= Set UV Lamp state (0 = off, 1 = on)


12.5

Profile Control Commands

Code

Meaning

Format

PC Clear

Profile

(clears profile
step)

PC<Space><Arg1><Space><Arg2><CR>
where Arg1= Profile number (range 1 to 4) and
Arg2=Step number (range 1 to 15)

NOTE: If Arg2 is not included, all 15 steps of the
selected profile will be cleared.

PM Profile

Mode

(returns status of
current profile
run/hold mode)

PM<CR>
There are no parameters to this command.

Return data format:
<Arg1><Space><Arg2><Space><Arg3><CR>
where Arg1=Run/hold status (0 = end, 1 = run),
Arg2=Profile number and
Arg3=Step number

NOTE: If a profile has not been selected to run, a 0
will be returned for Profile number and Step number.

...continued...








This manual is related to the following products: