beautypg.com

Warning – Eppendorf Innova 43R Incubator Shakers User Manual

Page 59

background image

59

New Brunswick

Operating manual

In the event the lid does not stay in the fully open position, you may want to adjust the lid
tension: tighten the 5/16-inch nuts on the hinges to increase spring tension.

9.8

Replacing Optional Germicidal Lamp

WARNING!

Before installing or removing the germicidal “UV” lamp, the authorized
technician will:
(1) Make sure power is completely disconnected from the shaker,
(2) Wear plastic gloves to protect the lamp from skin oils and to protect
your hands just in case the lamp should break,
(3) Before removal, make sure the lamp has fully cooled.


To replace the UV lamp, your service technician will follow all the WARNINGS above
and the procedure and WARNINGS outlined below, with reference to Figure 29:

1. Make sure power is disconnected from the shaker.
2. Open the Service Compartment by removing the side panel.
3. Free the heater cable harness and the temperature probe from their tie wraps and

Permagum

®

.

Figure 29: Germicidal (UV) Lamp Replacement

NOTE that wires, tie wraps and Permagum

®

are not shown above.

4. Remove the evaporator cover from the Service Compartment, paying careful attention

to the drain tubes.

This manual is related to the following products: