Table 4: Alarms
Indication
Description
TEMP
The temperature deviates more than 1ºC from setpoint after
achieving control temperature range. After lid is opened,
alarm will be disabled for 5 minutes while chamber recovers
to setpoint.
SPEED
The speed deviates more than 5 RPM from setpoint after
achieving operating speed setpoint. After lid is opened,
alarm will be disabled for 5 minutes while chamber recovers
to setpoint.
POWER
Indicates shaker is powering up (both at normal power-up
and after power interruption); will flash until the Control Knob
is moved.
HRS
Indicates when timed run is completed.
5.6 Lid
Access to the chamber of the Innova 43/43R is through the top-opening hinged lid. To
open the lid, firmly press the pedal or lift up on the front handle (see Figure 1). The lid
stands on its hinges until you choose to close it.
When the lid opens, the following will happen:
Heater turns off
Shaker stops
Interior light goes on and, when it is in
AUTO
mode, will remain on for 15
seconds after the lid is closed
UV germicidal lamp (if so equipped) turns off
5.7
Spill Pan/Water Reservoir
The Innova 43/43R is equipped with a spill pan and spill cover to protect the drive
mechanism in the case of accidental spills and/or broken glassware. This pan can also be
used as a water reservoir to humidify the chamber and to reduce evaporation. An
optional factory-installed humidity monitor is also available.
The reservoir can be drained through the quick-connect valve on the front of the shaker.