Figure 12a: humidity monitor (optional), Specifications, 1humidity monitor – Eppendorf Innova 43R Incubator Shakers User Manual
Page 28

28
Innova 43 & 43R Incubator Shakers
M1320-0050
Operating manual
Figure 12a: Humidity Monitor (optional)
1
Humidity monitor
When the humidity sensor is present in the chamber, maximum shaker temperature is
automatically limited to 60°C.
Specifications:
Humidity operating range:
0 to 100% RH
Accuracy:
See Figure 12b
Temperature operating range:
4 to 60ºC
Hysteresis:
± 1.5% RH
Recovery time after saturation: 10 seconds
Long term stability: 0.5% RH per year
Figure 12b: RH Sensor Operating Range & Humidity Error Graph
0
10
20
30
40
50
60
70
80
90
100
0
10
20
30
40
50
60
Temperature (C)
% R
H
+/- 5% Error
+/- 3% Error
+/- 5% Error
INTERMITTENT
OPERATION