Innova 43 & 43R Incubator Shakers

M1320-0050

Operating manual

Symptom(s)

Probable Cause(s) & Solution(s)






Photosynthetic GRO
lamp does not work.

Shaker is running in Program mode (look for Run icon in
display), and program calls for GRO lamp to be off.
GRO lamp mode has been changed by RS-232 command/
computer interface.
Bulb is burned out; replace (see Section 9.6).
Fuse is burned out: replace (see Section 9.4)
Check ballast voltage.
Check cable connections.
Check switch settings.
GRO lamp mode has been changed by RS-232
command/computer interface.

UV germicidal lamp
does not work.

Shaker is running in Program mode (look for Run icon in
display), and program calls for UV lamp to be off.
UV lamp is burned out; replace (see Section 0)
Fuse is burned out: replace (see Section 9.4)
Check ballast voltage.
Check cable connections.
Check switch settings.
UV lamp mode has been changed by RS-232 command/
computer interface.

9.2 Product

Returns

Should you need to return your Innova 43/43R to Eppendorf for any reason. Contact
Eppendorf Customer for return instructions.