Eppendorf Innova 43R Incubator Shakers User Manual

Page 17

background image

17

New Brunswick

Operating manual

Table 1: Screens

Screen

Name

Meaning

Features/Modes

DISP

Display

Shows two user-selectable parameters* and actual values.

SUMM

Summary

Shows all parameters*, setpoints and actual values.

SET

Set-Up

Set day of week, set time, enable or mute alarm, lock or
unlock operating parameters

LAMP Lamps

Internal chamber light: ON (always on); OFF (always
off); AUTO (default mode), light goes on and stays on
when lid is open, shuts off 15 seconds after lid is closed,
and goes on for 15 seconds when Control Knob is moved.
Photosynthetic lights (GRO): ON, OFF,

NONE

UV Light (UV): ON, OFF,

NONE

COMM Communication

(RS-232)

SET: set baud rate
OFF disables RS-232
MONITOR: PC commands shaker to read setpoints and
actual values on a schedule determined by PC software.
Parameters are unlocked and can be changed by program
or manually.
SLAVE: PC controls shaker and logs data.
TALK: Shaker sends setpoint and actual data to PC at
one-minute intervals.

CAL

Calibration

Allows user to enter a temperature offset.
Self-calibrates the speed sensor.

PROG

Program

Allows user to set up 1- 4 programs, each with 1-15 steps.

* see Table 2, Display Screen Parameters

 Optional

Not installed


Table 2: Display Screen Parameters

Parameter Name

Meaning

RPM

Shaking speed, in revolutions/minute

ºC

Chamber temperature, in degrees Celsius

HRS

Programmed time remaining, in hours

 %RH

Relative Humidity, in percent.

 UV

Status of Ultraviolet germicidal lamp

 GRO

Status of Photosynthetic growth lamps

 Optional

5.4 Display

Icons

There are six icons that help identify operating conditions. Five of these icons are
located at the bottom of the display, and they are visible, when applicable, in any screen
you are viewing, except the Program subscreens. The sixth appears, when applicable,
next to ºC whenever the temperature is onscreen.