beautypg.com

Warning – Eppendorf Innova 43R Incubator Shakers User Manual

Page 58

background image

58

Innova 43 & 43R Incubator Shakers

M1320-0050

Operating manual

3. Remove the two thumbscrews from the diffuser holder.

Figure 28: Replacing Photosynthetic Lamps

4. Remove the diffuser by placing a finger underneath and rotating the diffuser in the

slot at the back of the photosynthetic light assembly. All 9 lamps should now be
accessible to you.

5. To remove the lamp for replacement, rotate the lamp 90º, taking care to support the

lamp before it drops down.

6. To replace the lamp, orient the pins of the new lamp vertically. Simultaneously,

gently press both sides of the lamp into the slots.

7. Rotate the new lamp 90º. The lamp should catch in the holders on both ends. Make

sure both ends are secure.

8. Put the diffuser back in place, reinstalling the diffuser holder with its two

thumbscrews.

9.7

Adjusting Lid Tension

WARNING!

The following procedures are provided for your information only.
Do not attempt to perform these service interventions yourself unless you
are an authorized service technician.