Exit and save calibration – Class1 FoamLogix 2.1A & 1.7AHP REV D User Manual
Page 57
FoamLogix, Model 2.1A and 1.7AHP Class “A”
Electronic Foam Proportioning Systems
57
4. Start the Hale FoamLogix foam pump by pressing
the red
ON button.
The LEDs on the horizontal bar graph lights as
the foam pump begins operating at approximately
two-thirds speed, pump ing foam concentrate into
the container.
The display shows the volume of foam
concentrate being pumped.
5. STOP the foam pump and accurately measure
the amount of foam concentrate collected.
6. Adjust the reading on the display to match the
volume actually pumped using the or
button.
7. Repeat the procedure to verify the setting is
correct.
8. Set the bypass valve handle back to
INJECT
position.
Record Foam Pump Feedback Calibration
Factor
1. Press and release the
i button.
The display shows
FA xxx,the foam pump
feedback calibration factor.
(See Figure 33:
“Display - Foam Pump Feedback Calibration
Factor.”)
2. Record this value for future reference. This
factor must be programmed into the display if
the display is ever replaced.
FOAM PUMP FEEDBACK CALIBRATION
FACTOR: __________________________
EXIT AND SAVE CALIBRATION
1. To exit calibration and save the set values,
press and hold the
i button.
Figure 33: Display - Foam Pump Feedback Calibration
Factor
The display show
PASS, then clears.
2. While continuing to hold the display button
enter the password ().
3. The display shows
SCAL for several seconds and
then cycles through the start-up sequence
followed by the flow display
O. (See Figure 34:
“Display - Exit and Save Calibration.”)
This completes verification and adjustment of the
system. The Hale FoamLogix system is now ready
to be placed in service.
FOAMLOGIX, Model 2.1A Class "A"
55
Electronic Foam Proportioning System
Installation
ISO 9001 CERTIFIED
4. Start the Hale FoamLogix foam pump by
pressing the red
ON button.
The LEDs on the horizontal bar graph
lights as the foam pump begins operating
at approximately two-thirds speed, pump-
ing foam concentrate into the container.
The display shows the volume of foam
concentrate being pumped.
5. STOP the foam pump and accurately
measure the amount of foam concentrate
collected.
6.
Adjust the reading on the display to match
the volume actually pumped using the
or button.
7. Repeat the procedure to verify the setting
is correct.
8. Set the bypass valve handle back to
INJECT position.
Record Foam Pump Feedback
Calibration Factor
1. Press and release the
i button.
The display shows
FA xxx, the foam pump
feedback calibration factor. (See Figure
33: “Display - Foam Pump Feedback
Calibration Factor.”)
2. Record this value for future reference.
This factor must be programmed into the
display if the display is ever replaced.
FOAM PUMP FEEDBACK CALIBRATION
FACTOR: __________________________
EXIT AND SAVE CALIBRATION
1. To exit calibration and save the set values,
press and hold the
i button.
The display show
PASS, then clears.
2. While continuing to hold the display button
enter the password ().
3. The display shows
SCAL for several
seconds then cycles through the start-up
sequence followed by the flow display
O.
(See Figure 34: “Display - Exit and Save
Calibration.”)
This completes verification and adjustment of the
system. The Hale FoamLogix system is now
ready to be placed in service.
Figure 33: Display - Foam Pump Feedback Calibration
Factor
Press and
release - the
feedback factor
is shown
Press and
release -
SCAL
is shown
Figure 34: Display - Exit and Save Calibration
Figure 34: Display - Exit and Save Calibration
FOAMLOGIX, Model 2.1A Class "A"
55
Electronic Foam Proportioning System
Installation
ISO 9001 CERTIFIED
4. Start the Hale FoamLogix foam pump by
pressing the red
ON button.
The LEDs on the horizontal bar graph
lights as the foam pump begins operating
at approximately two-thirds speed, pump-
ing foam concentrate into the container.
The display shows the volume of foam
concentrate being pumped.
5. STOP the foam pump and accurately
measure the amount of foam concentrate
collected.
6.
Adjust the reading on the display to match
the volume actually pumped using the
or button.
7. Repeat the procedure to verify the setting
is correct.
8. Set the bypass valve handle back to
INJECT position.
Record Foam Pump Feedback
Calibration Factor
1. Press and release the
i button.
The display shows
FA xxx, the foam pump
feedback calibration factor. (See Figure
33: “Display - Foam Pump Feedback
Calibration Factor.”)
2. Record this value for future reference.
This factor must be programmed into the
display if the display is ever replaced.
FOAM PUMP FEEDBACK CALIBRATION
FACTOR: __________________________
EXIT AND SAVE CALIBRATION
1. To exit calibration and save the set values,
press and hold the
i button.
The display show
PASS, then clears.
2. While continuing to hold the display button
enter the password ().
3. The display shows
SCAL for several
seconds then cycles through the start-up
sequence followed by the flow display
O.
(See Figure 34: “Display - Exit and Save
Calibration.”)
This completes verification and adjustment of the
system. The Hale FoamLogix system is now
ready to be placed in service.
Figure 33: Display - Foam Pump Feedback Calibration
Factor
Press and
release - the
feedback factor
is shown
Press and
release -
SCAL
is shown
Figure 34: Display - Exit and Save Calibration