Simulated flow, Foam concentrate injection rate – Class1 FoamLogix 2.1A & 1.7AHP REV D User Manual
Page 55

FoamLogix, Model 2.1A and 1.7AHP Class “A”
Electronic Foam Proportioning Systems
55
Press the
3.
or button and set the reading to
match the actual flow calculated from the Pitot
gauge reading.
Decrease fire pump pressure by approxi mately
4.
one half (1/2) and recalculate water flow rate.
Verify the reading on the control unit is within 5%
of the actual value.
STOP the water flow when adjustments are
5.
completed.
Record Flow and Sensor Calibration Factors
Press and release the
i button. The display show
F xx.x, which is the water flow sensor calibration
factor (See Figure 28: “Display Flow Sensor
Calibration Factor.”)
FOAMLOGIX, Model 2.1A Class "A"
53
Electronic Foam Proportioning System
Installation
ISO 9001 CERTIFIED
3. Press the
or button and set the
reading to match the actual flow calculated
from the Pitot gauge reading.
4. Decrease fire pump pressure by approxi-
mately one half (1/2) and recalculate water
flow rate.
Verify the reading on the control unit is
within 5% of the actual value.
5. STOP the water flow when adjustments
are completed.
Record Flow and Sensor Calibration
Factors
Press and release the
i
button. The display
show
F xx.x, which is the water flow sensor
calibration factor (See Figure 28: “Display -
Flow Sensor Calibration Factor.”)
Record this value for future reference. This
factor must be programmed into the display if
the display is ever replaced.
WATER FLOW SENSOR CALIBRATION
FACTOR: __________________________
SIMULATED FLOW
The default Simulated Flow value is factory set to
150 GPM (568 LPM) and, if necessary, may be
adjusted while in the user calibration mode.
Press the
i
button. The default simulated flow
rate is shown. (See Figure 29: “Display - Simu-
lated Flow Calibration.”)
Adjust the setting to the required rate by pressing
the or buttons (e.g.,
S 150).
FOAM CONCENTRATE INJECTION
RATE
When the Hale FoamLogix system power is
turned ON, the foam concentrate injection rate
stored in memory is the default setting. The user
specific default concentrate injection rate is
adjusted in calibration mode.
1. Press the
i button.
The display shows the current default
concentrate injection rate stored in the
computer memory for the selected foam
concentrate tank. (See Figure 30: “Display
- Foam Concentrate Injection Rate Default
Value” on page 52.)
Figure 28: Display - Flow Sensor Calibration Factor
Press and
release -
F xx.x
is shown
Figure 29: Display - Simulated Flow Calibration
Press and
release -
CAL is
shown
Figure 28: Display - Flow Sensor Calibration Factor
Record this value for future reference. This factor
must be programmed into the display if the display is
ever replaced.
WATER FLOW SENSOR CALIBRATION
FACTOR: __________________________
SIMULATED FLOW
The default Simulated Flow value is factory set to
150 GPM (568 LPM) and, if necessary, may be
adjusted while in the user calibration mode.
FOAMLOGIX, Model 2.1A Class "A"
53
Electronic Foam Proportioning System
Installation
ISO 9001 CERTIFIED
3. Press the
or button and set the
reading to match the actual flow calculated
from the Pitot gauge reading.
4. Decrease fire pump pressure by approxi-
mately one half (1/2) and recalculate water
flow rate.
Verify the reading on the control unit is
within 5% of the actual value.
5. STOP the water flow when adjustments
are completed.
Record Flow and Sensor Calibration
Factors
Press and release the
i
button. The display
show
F xx.x, which is the water flow sensor
calibration factor (See Figure 28: “Display -
Flow Sensor Calibration Factor.”)
Record this value for future reference. This
factor must be programmed into the display if
the display is ever replaced.
WATER FLOW SENSOR CALIBRATION
FACTOR: __________________________
SIMULATED FLOW
The default Simulated Flow value is factory set to
150 GPM (568 LPM) and, if necessary, may be
adjusted while in the user calibration mode.
Press the
i
button. The default simulated flow
rate is shown. (See Figure 29: “Display - Simu-
lated Flow Calibration.”)
Adjust the setting to the required rate by pressing
the or buttons (e.g.,
S 150).
FOAM CONCENTRATE INJECTION
RATE
When the Hale FoamLogix system power is
turned ON, the foam concentrate injection rate
stored in memory is the default setting. The user
specific default concentrate injection rate is
adjusted in calibration mode.
1. Press the
i button.
The display shows the current default
concentrate injection rate stored in the
computer memory for the selected foam
concentrate tank. (See Figure 30: “Display
- Foam Concentrate Injection Rate Default
Value” on page 52.)
Figure 28: Display - Flow Sensor Calibration Factor
Press and
release -
F xx.x
is shown
Figure 29: Display - Simulated Flow Calibration
Press and
release -
CAL is
shown
Figure 29: Display - Simulated Flow Calibration
Press the
i button. The default simulated flow rate
is shown. (See Figure 29: “Display - Simu lated Flow
Calibration.”)
Adjust the setting to the required rate by pressing
the or buttons (e.g.,
S 150).
FOAM CONCENTRATE INJECTION RATE
When the Hale FoamLogix system power is turned
ON, the foam concentrate injection rate stored in
memory is the default setting. The user specific
default concentrate injection rate is adjusted in
calibration mode.
1. Press the
i button.
The display shows the current default concentrate
injection rate stored in the computer memory for the
selected foam concentrate tank. (See Figure 30:
“Display Foam Concentrate Injection Rate Default
Value” on page 56.)