Foam pump feedback calibration – Class1 FoamLogix 2.1A & 1.7AHP REV D User Manual
Page 56
FoamLogix, Model 2.1A and 1.7AHP Class “A”
Electronic Foam Proportioning Systems
56
54
FOAMLOGIX, Model 2.1A Class "A"
Electronic Foam Proportioning System
Installation
ISO 9001 CERTIFIED
2. If the factory default values have not been
changed the display shows
PA 0.5.
3. Use the or buttons to set the user
specified default concentrate injection rate.
FOAM PUMP FEEDBACK
CALIBRATION
IMPORTANT!
FOAM PUMP FEEDBACK IS CALIBRATED AFTER
INSTALLATION TO VERIFY VALUES WITH THE
ACTUAL FOAM CONCENTRATE(S) BEING USED.
ONLY CALIBRATE USING ACTUAL FOAM CON-
CENTRATES.
DO NOT USE WATER, TRAINING OR TEST FOAMS
FOR FEEDBACK CALIBRATION VERIFICATION.
1. Press the
i button.
The display shows
A x.xx, the total vol-
ume of foam concentrate pumped during
the last calibration run. (See Figure 31:
“Display - Foam Pump Feedback Calibra-
tion.”)
2. Set the bypass valve to the
BYPASS.
3. Place a graduated measure container at
the outlet of the bypass hose capable of
containing the expected volume of foam
concentrate, 5 gallons (19 liters) minimum.
(See Figure 32: “Foam Concentrate
Collection.”).
Note: If an accurate calibrated container is
not available an accurate scale can be used
to weigh the foam concentrate pumped. The
total volume of foam concentrate is then
calculated from this weight and the density of
the foam concentrate per the MSDS sheet.
Figure 30: Display - Foam Concentrate Injection Rate
Default Value
Press and
release - the
default value is
shown
Figure 32: Foam Concentrate Collection
Measuring
(calibrated)
Container
Bypass Valve on
Foam Pump
Bypass Hose
Figure 31: Display - Foam Pump Feedback Calibration
Press and
release - the
present value is
shown
Figure 30: Display - Foam Concentrate Injection Rate
Default Value
2
.
If the factory default values have not been
changed the display shows
PA 0.5.
3. Use the or buttons to set the user specified
default concentrate injection rate.
FOAM PUMP FEEDBACK CALIBRATION
IMPORTANT!
FOAM PUMP FEEDBACK IS CALIBRATED AFTER
INSTALLATION TO VERIFY VALUES WITH THE
ACTUAL FOAM CONCENTRATE(S) BEING USED.
ONLY CALIBRATE USING ACTUAL FOAM CON-
CENTRATES.
DO NOT USE WATER, TRAINING OR TEST FOAMS
FOR FEEDBACK CALIBRATION VERIFICATION.
1. Press the
i button.
The display shows
A x.xx, the total vol ume of foam
concentrate pumped during the last calibration run.
(See Figure 31: “Display - Foam Pump Feedback
Calibra tion.”)
2. Set the bypass valve to the
BYPASS.
54
FOAMLOGIX, Model 2.1A Class "A"
Electronic Foam Proportioning System
Installation
ISO 9001 CERTIFIED
2. If the factory default values have not been
changed the display shows
PA 0.5.
3. Use the or buttons to set the user
specified default concentrate injection rate.
FOAM PUMP FEEDBACK
CALIBRATION
IMPORTANT!
FOAM PUMP FEEDBACK IS CALIBRATED AFTER
INSTALLATION TO VERIFY VALUES WITH THE
ACTUAL FOAM CONCENTRATE(S) BEING USED.
ONLY CALIBRATE USING ACTUAL FOAM CON-
CENTRATES.
DO NOT USE WATER, TRAINING OR TEST FOAMS
FOR FEEDBACK CALIBRATION VERIFICATION.
1. Press the
i button.
The display shows
A x.xx, the total vol-
ume of foam concentrate pumped during
the last calibration run. (See Figure 31:
“Display - Foam Pump Feedback Calibra-
tion.”)
2. Set the bypass valve to the
BYPASS.
3. Place a graduated measure container at
the outlet of the bypass hose capable of
containing the expected volume of foam
concentrate, 5 gallons (19 liters) minimum.
(See Figure 32: “Foam Concentrate
Collection.”).
Note: If an accurate calibrated container is
not available an accurate scale can be used
to weigh the foam concentrate pumped. The
total volume of foam concentrate is then
calculated from this weight and the density of
the foam concentrate per the MSDS sheet.
Figure 30: Display - Foam Concentrate Injection Rate
Default Value
Press and
release - the
default value is
shown
Figure 32: Foam Concentrate Collection
Measuring
(calibrated)
Container
Bypass Valve on
Foam Pump
Bypass Hose
Figure 31: Display - Foam Pump Feedback Calibration
Press and
release - the
present value is
shown
Figure 31: Display - Foam Pump Feedback Calibration
3.
Place a graduated measure container at the
outlet of the bypass hose capable of containing
the expected volume of foam concentrate, 5
gallons (19 liters) minimum.
(See Figure 32: “Foam Concentrate Collection.”)
Note: If an accurate calibrated container is not
available an accurate scale can be used to
weigh the foam concentrate pumped. The total
volume of foam concentrate is then calculated
from this weight and the density of the foam
concentrate per the MSDS sheet.
Bypass Hose
Bypass Valve
on Foam
Pump
Figure 32: Foam Concentrate Collection