RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 89
- 89 -
The display indicates initial mass of tested sample. In order to ensure stable
measurement conditions, it is recommended to close the drying chamber’s lid
before tarring the indication.
Press
key to tare drying pan mass, and save its mass in moisture analyzer’s
memory.
Mass indication returns to ZERO and the following command is displayed.
If
activates a bar graph indicating current sample mass in relation to saved Min
and Max limits for sample’s mass.
Load tested sample onto the drying pan. Select sample’s mass according to
previously carried out tests and characteristics of tested substance / material.
This manual is related to the following products: