An authorized RADWAG representative determines the minimum sample

dependent on packaging mass. The process is carried out with mass

standards, in balance place of use and in accordance with the requirements

of the accepted quality system. The obtained values are entered to balance

software: .

A minimum sample determining method may have a few tare values

defined, along with assigned values of minimum samples, expiry dates of

the carried out measurements and entered data. The settings are not to be

edited by balance user.

Using mode guarantees that the results of completed

weighing processes are within set tolerance limits, and they comply with

the requirements of the quality system accepted by an organization.

CAUTION: the mode operates in the weighing mode only.

Available options:

METHOD

It is a means for designating an accepted quality standard.

Pressing a field displays a window with a list of minimal sample

weight values stored in balance’s memory, following a criterion on

determining the minimum sample.

Adding a new method is available only from the menu level

MODE

Lock – on selecting this option and during the weighing process

the display indicates corresponding pictograms that inform a user

whether the weighed mass is below or above the minimum sample

limit. Additionally the software disables accepting the measurement

result if it is below the minimum sample limit

Warn – on selecting this option and during the weighing process

the display indicates corresponding pictograms that inform a user

whether the weighed mass is below or above the minimum sample

limit. The user can accept a measurement that is below the

minimum sample limit, but on a printout such measurement is

preceded by an asterisk (*).

CAUTION:

Means of setting a method of minimum sample determining is described

in point 27.9. of this user manual: .

RADWAG authorized personnel is only authorized to add new values of

minimum samples.

Weighing with application of mode

If during a weighing process a user wants to be informed whether a

measurement exceeds the minimum sample limit for a specific weighing

This manual is related to the following products: