Manual printout / automatic printout – RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 144
- 144 -
n – parameter, decimal value specifying working mode number. See OMI
command description for detailed information.
Example
Command:
OMG
– give current working mode
Response:
OMG_13_OK
– the device currently operates in Statistics mode
Send all implemented commands
Format: PC CR LF
Accessible responses:
PC_A_”Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,
LOGIN,LOGOUT,PC" - command carried out, terminal has sent all
implemented commands.
27.3. Manual printout / Automatic printout
Moisture analyzer MA 3Y series enables generating manual or automatic
printouts.
•
Manual printout: press
key on stabilization of indication
(measurement result),
•
Automatic printout is generated automatically, in accordance to the settings
of automatic printout (see point 11.6 of this user manual).
The content of a printout depends on setting in menu
Format of mass printout:
1
2
3
4 -12
13
14 15 16 17 18
stability
marker
space character mass space
unit
CR LF
Stability marker
[space] if stable
[?] if unstable
[!] air buoyancy correction mode is enabled
[^] if error of exceeding range to + occurs
[v] if error of exceeding range to – occurs
Character
[space] for positive values
[-] for negative values
Mass
9 characters with decimal point and right justification
Unit
3 characters with left justification