12.4. Drying chamber test
Parameter
indicated by the moisture analyzer’s internal thermometer.
The testing procedure is carried out using a dedicated kit, available as optional
equipment of the moisture analyzer. Means of assembling the testing kit is
described in the previous point (temperature adjustment).
In order to start the testing procedure assemble the testing kit inside the drying
chamber, and next go to menu group
pressing
A new window is opened for
entering temperature value set for
checking the moisture analyzer.
Then enter a permissible error for
the set temperature value (the
error cannot be lower than the
default factory value). Next enter
the factory no. of the temperature
adjustment kit, and press
field. The software orders heating
the drying chamber. Control time
for a given temperature value is 8
minutes.
During the heating time the
moisture analyzer displays a
window with current data on
temperature inside the drying
chamber and heating time.