Proximity sensors, Information, Printouts – RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 62: Information 16.5. printouts
- 62 -
16.3.1. Proximity sensors
The balance comprises two proximity sensors enabling wave control of the
balance therefore the user does not have to either press any key on the
overlay or touch the screen.
The program recognizes 4 types of motions performed near the sensor:
1. Placing hand near the left sensor
2. Placing hand near the right sensor
3. Motion to the left
4. Motion to the right
CAUTION: using option LEFT MOTION or RIGHT MOTION, it is necessary to
remember that
500 ms!
To each of the motions, one of the available options may be assigned:
[None; Profile; Adjustment; Zeroing; Tarring; Tare setting; Tare switch off;
Reset tare; Packaging; Print; Header printing; Footer printing; Unit;
Universal Variable 1; Universal Variable 2; Universal Variable 3; Universal
Variable 4; Universal Variable 5; Confirm; Abort; User; Left door; Right
door; Open/Close door; Parameters; Product, Warehouse; Client].
The selected option is recognized by the program and performed when
activated by a particular motion.
In order to ensure correct operation the user must remember to set the
right sensors sensitivity and the right value for proximity sensors delay (for
more information refer to point 33of this manual).
16.4. Information
Information on current process is
displayed in the grey coloured workspace.
It contains maximally 4 pieces of
information. If larger amount of data is
selected, then only the first 4 are
displayed. Each parameter features two
option:
- YES, information enabled (displayed in
the workspace)
- NO, information disabled
16.5. Printouts
Menu printouts comprises two separate sections. The first one is standard
printouts, the other one is non-standard printouts.
Standard printout
comprises three internal sections featuring different variables. For each
variable set option YES to include it in a standard printout. If option NO is