Transport and storage, Delivery check, Packaging – RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 14: Unpacking, assembling and startup, Assembling and place of use, Delivery check 3.2. packaging

background image

- 14 -

3. TRANSPORT AND STORAGE

3.1. Delivery check

Please check the packaging immediately upon delivery and check instrument’s
condition directly upon unpacking for any visible signs of external damage.

3.2. Packaging

Please retain all parts of the original packaging should the moisture analyzer
be transported in the future. Only the original packaging should be used for
dispatching the moisture analyzer. Before packing, disconnect all attached
cables and remove any loose/movable parts (weighing pan, shields, inserts).
Place moisture analyzer and its components in their original packaging, and
protect them against damage during transport.

4. UNPACKING, ASSEMBLING AND STARTUP

4.1. Assembling and place of use

A moisture analyzer should be stored and used in locations free of

vibrations and shakes, free of air movement and dust, located at the max
altitude of 2000 above sea level,

Moisture analyzer’s place of use should ensure good air circulation around

the instrument (approximately 20 cm fee space around the moisture

analyzer and 1 m free space over it),

Ambient air temperature in the weighing room should not exceed the

range

of:

+10 °C ÷ +40 °C,

Ambient relative humidity should not exceed 80% in the temperature up to

31°C, and decrease linearly to 50 % of relative humidity in temperature
40°C,

A moisture analyzer should be located on a stable wall console desk or a

stable working table which is not affected by vibrations and distant from
heat sources,

Take special safety measures when weighing magnetic objects, as part of

the balance is a strong magnet.