15. PROFILES
A Profile is a data pack determining:
o
functioning of working modes, e.g. weighing and drying,
o
type of data displayed during working mode operation,
o
active function keys,
o
accessible set of measuring units
o
mandatory criteria for moisture analyzer’s speed of operation and
measurement stability,
The software enables creating numerous profiles, which in practice provides:
o
each user to create their own and individual operating environment
o
moisture analyzer operation can be easily programmed by activating used
defined function keys and information on process (improving ergonomics
of operation)
15.1. Creating a profile
A default profile is named
by:
−
Copying an already existing profile and its modification
−
Creating a new profile
Copying an existing profile
Procedure:
•
Enter moisture analyzer’s main menu by pressing Setup key,
•
Enter submenu < Profiles>,
•
Press and hold an entry with profile name that should be copied,
•
A menu is opened referring to available options:
o
Edit
o
Delete
o
Copy
o
Cancel
•
Select option
•
A new profile named
same as in the base profile,
•
After copying change necessary data in a profile: (name, etc.)
Creating a new profile
Procedure:
•
Enter moisture analyzer’s main menu by pressing Setup key,
•
Enter submenu < Profiles>,
•
Press
key, which opens a message box:
•
Accept the message box by pressing
key. The software automatically
adds a new record and enters its editing mode.