On completing the adjustment procedure the moisture analyzer indicates
a message box:
•
On confirming the message box with
key, the moisture analyzer
returns to the main mode
12.2. User adjustment
User adjustment is carried out with an optional standard, which mass ranges
between 0,15 Max ÷ Max capacity. Adjustment procedure is compatible with
the external adjustment process, but before its start the software opens a
message box for entering mass of a standard used for user adjustment
procedure.
In order to start user adjustment, enter submenu < Adjustment >, and select
option: “ User adjustment”. Then follow the commands indicated on the
display.
12.3. Temperature adjustment
Temperature adjustment is carried out with a dedicated kit, that is an optional
equipment of the moisture analyzer MA 3Y series.
1. Thermometer
2. Thermometer holder and drying chamber shield
Adjustment of the drying chamber temperature is a process carried out for the
purpose of adjusting the temperature sensor assembled inside the drying