beautypg.com

RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 46

background image

- 46 -

After 8 minutes, the system

opens a window with an on-

screen numeric keyboard.

Using the on-screen keyboard

insert the current temperature

value of the moisture analyzer,

that is indicated on the

external control thermometer.

Press key to proceed

to the following adjustment

step.

The moisture analyzer

proceeds to the last step of

adjustment process. It heats

the drying chamber to a higher

temperature and for another 8

minutes it maintains this

temperature and displays its

value.

After 8 minutes, the system

opens a window with an on-

screen numeric keyboard.

Using the on-screen keyboard

insert the current temperature

value of the moisture analyzer,

that is indicated on the

external control thermometer.

Press key to proceed

to the following adjustment

step to complete temperature

adjustment procedure.

On completing temperature adjustment procedure, the moisture analyzer

displays a corresponding message box. Accept the message box by pressing

key and return to the main window.