Ambient conditions – RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 104

- 104 -
ranging from 0.0001 g to 10.0000 g
(
3
50.0000 g
2.5000 g Minimal sample weight for all net masses
being weighed in packaging of mass
ranging from 10.0001 g to 50.0000 g
(
4
200.0000 g 4.0000 g Minimal sample weight for all net masses
being weighed in packaging of mass
ranging from 50.0001 g to 200.0000 g
(
Example no. 2 for AS 220.3Y:
No. Tare value
Minimal
sample
weight
Description
1
220.0000 g 0.5000 g Minimal sample weight for all net masses
being weighed either in packaging of any
mass of full permissible range (
button in use) or without packaging
(
Example no. 3 for AS 220.3Y:
No. Tare value
Minimal
sample
weight
Description
1
0.0000 g
0.2500 g Minimal sample weight for all net masses
being weighed without packaging (
button not in use).
Program identifies setting for which
minimal sample weight is valid only for
weighing samples without packaging. Upon
the use of tarring option, the icon
informing about minimal sample weight
being in use turns off. For weighing with
tare no minimal sample weight is defined.
The user can view the entered data but it cannot be edited.
19.9. Ambient conditions
Database of ambient conditions contains parameters related to measurement
of ambient conditions at a workstation. Depending on moisture analyzer’s
configuration, a report may contain data on temperature, humidity and value
of atmospheric pressure. In case a THB ambient conditions module is
connected to a moisture analyzer, records from the module are also saved.
Procedure
•
Enter submenu < Databases>, and press <
Ambient conditions >
field