RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 45

- 45 -
Before starting the temperature adjustment process, close drying chamber lid
and follow below process descriptions:
Enter submenu group
Start adjustment> application by pressing field with its name. The moisture analyzer displays a message box on procedure start, and data on the process (step number, current temperature value, heating time, burn-in (heating) time). After 8 minutes, the system opens a window with an on- screen numeric keyboard. Using the on-screen keyboard insert the current temperature value of the moisture analyzer, that is indicated on the external control thermometer. Press to the following adjustment step. The moisture analyzer switches on the halogen heater and for another 8 minutes it heats the drying chamber and simultaneously displays current temperature value and other parameters.