RADWAG MA 200.3Y Moisture Analyzer User Manual

Page 45

background image

- 45 -

Before starting the temperature adjustment process, close drying chamber lid

and follow below process descriptions:

Enter submenu group

.

Start

adjustment> application by

pressing field with its name.

The moisture analyzer displays

a message box on procedure

start, and data on the process

(step number, current

temperature value, heating

time, burn-in (heating) time).

After 8 minutes, the system

opens a window with an on-

screen numeric keyboard.

Using the on-screen keyboard

insert the current temperature

value of the moisture analyzer,

that is indicated on the

external control thermometer.

Press key to proceed

to the following adjustment

step.

The moisture analyzer switches

on the halogen heater and for

another 8 minutes it heats the

drying chamber and

simultaneously displays current

temperature value and other

parameters.