23.14. Start logo

Caution:

This parameter is available only for authorized operator of a moisture analyzer.

The parameter enables changing the image displayed on the terminal on

moisture analyzer start up.

23.15. Export system events

Caution:

This parameter is available only for authorized operator of a moisture analyzer.

The parameter is used to generating a special type of a file that is saved

automatically on parameter enabling on an external data storage device plugged

to moisture analyzer’s USB port.

The file is used by RADWAG service personnel to diagnose potential problems in

moisture analyzer operation.

Procedure:

Plug an external data storage device to the USB port

Enter submenu <

Others>

Select parameter ,

The software automatically generates a file and saves it on the external

data storage device.

Send the generated file to the RADWAG company.

24. UPDATE

The function contains two modules which enable updating:

o

Area of a user: SOFTWARE

o

Metrological parameters: WEIGHING MODULE.

Updating process takes place automatically by uploading data from a data

storage device connected to moisture analyzer’s USB port.

Software updating procedure:

o

Prepare a data storage device with update file. Required file extension:

*.pue7

o

Connect the data storage device to moisture analyzer’s USB port

o

Select menu

o

The display previews content saved on the data storage device, search

for the update file and press its name.

o

Updating process takes place automatically