Minimum sample weight – RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 37

- 37 -
Never – printout deactivated
First stable
– the first stable measurement result obtained after
placing a load on moisture analyzer’s weighing pan is
recorded, record of the following measurement result
takes place only on unloading the weighing pan,
returning of the mass indication below set
threshold’s value, and placing another load on
balance’s weighing pan
Last stable
– accepted is the last stable measurement recorded
before taking the load off the weighing pan. The
record takes place on taking the load off the
balance’s weighing pan, and returning of the mass
indication below set threshold’s value.
THRESHOLD – mass value obligatory for operation of automatic printout.
Set in grams.
−
Printout
Contains type of printout that is related to a working mode. Printout takes
place on pressing PRINT key on moisture analyzer’s overlay.
Accessible options:
•
STANDARD PRINTOUT
Possible declaration of printout content: HEADER, WEIGHING DATA
and FOOTER.
printout components marked as
pressing printout activating function key.
•
NON-STANDARD PRINTOUT
Database of printouts enables selecting one of available non-
standard printouts visible in menu
unique printout which is automatically added to the databases.
Caution:
Means of designing printouts is described in point 16.5 of this user
manual.
11.7. Minimum Sample Weight
The weighing mode settings comprise a function of
The function is enabled on specifying in the database sample > parameters on methods for determining the minimum sample and the values of minimal sample weight specific for a method. In standard version, this database item is not filled in with data. Activities related to determining minimum sample and adding data to the database: only for an authorized “RADWAG” personnel. Should balance operator use the minimum sample mode, and data on minimal sample weight are not entered to balance menu, then please contact the local representative of “RADWAG” company.