Drying temperature reducing, Compatibility with mac/max series – RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 127
- 127 -
Caution:
Means of leveling the moisture analyzer is described in point 11.3 of this user
manual.
23.9. Drying temperature reducing
This parameter automatically determines reducing the drying temperature after
a time interval specified for a moisture analyzer with Max drying temperature of
250 °C.
Means of operation:
While drying temperature holds within a limit between 161 °C and 250 °C then
the drying temperature can be maintained proportionally to set temperature
value, i.e. 1 hour for the temperature 161 °C and 20 minutes for the
temperature 250 °C.
While drying a sample in a temperature set to 250 °C, this temperature can be
maintained for the time of 20 minutes, and afterwards the software
automatically reduces the temperature to 160 °C (drying process is continued
without break).
The time of reducing the temperature from 250 °C to 160 °C is 20 minutes.
For the QUICK MODE the initial exceeding of the Max drying temperature is 30
%, but not higher than the maximum drying temperature of a moisture analyzer.
Procedure:
•
Enter submenu <
Others>
•
Select parameter
window with selectable values
•
Select one of the available options:
YES – temperature reduction mode enabled
NO – temperature reduction mode disabled
Selecting a value causes returning to the menu window.
Caution:
Remember that prolonged drying a sample is temperature above 160
°
C may
cause damaging the components of the moisture analyzer.
23.10. Compatibility with MAC/MAX series
This parameter enables reflecting operation of moisture analyzers MAC and MAX
series. By enabling the parameter (i.e. settings its value to ,YES>), the moisture
analyzer MA 3Y series starts functioning (in the field of drying temperatures and
control of the drying process) as the moisture analyzers MAC and MAX series (in
comparable temperature conditions).
The parameter enables a user to use the drying mode settings as in case of the
moisture analyzers MAC and MAX series. Thus, on purchasing a moisture