FILTER
Each measurement signal before being displayed is electronically
processed for the purpose of obtaining corrected parameters
specific for stable measurement result, i.e. ready to read.
In a specific extend a user can influence the range of signal processing by
selecting an appropriate FILTER value. Available options:
o
very fast
o
fast
o
average / normal
o
slow
o
very slow
While setting the filtering level the user should consider the actual
operating conditions of a moisture analyzer. In case of very good
operating conditions set the filter to average/normal or fast, in case of
rough conditions set the filter to slow or very slow.
Value release
This parameter determines displaying the stability marker for a
measurement result.
There are 3 available settings of value release parameter
o
fast
o
fast + reliable /recommended/
o
reliable
Caution:
The speed of stability marker occurrence depends on applied filter value and
applied value release parameter
Autozero function
The function is to automatically monitor and correct zero
indication.
If the function is enabled, the following measurement results are
compared to each other in constant time intervals. If the results differ
less than declared AUTOZERO range, e.g. 1 division, the moisture
analyzer is automatically zeroed, and the markers of stable indication
and precise zero
are displayed.
Active AUTOZERO means, that each measurement starts from the precise
zero point. There are, however, cases where the function may disturb the
measuring process, for instance during very slow load placing or pouring
onto the moisture analyzer’s weighing pan. In such case, the zero
indication correcting system may also correct the actual indication of a
load placed on the weighing pan.
Accessible settings: NO - autozero function disabled
YES - autozero function enabled