Drying time – RADWAG MA 200.3Y Moisture Analyzer User Manual
Page 71

- 71 -
place at full power of the filaments, causing heating the drying chamber to
increased temperature (maximum temperature value is higher by 30 % from
set drying temperature). Overheating compensates heat loss during
evaporation of large moisture content in the test’s initial stage.
STEP mode
Enables determining up to three steps with
random drying temperature. Step mode is
recommended for drying substances with
moisture content over 15%. Temperature and
heating time in each step have to be adjusted
by tests.
17.2.5. Drying time
Drying time is set by selecting one of available finish mode criteria. This
means that a moisture analyzer must meet pre-set criterion (one of the
requirements: mass over time, time) to automatically finish the drying mode.
Finish of drying process can be triggered by amount of time, independently on
loss in mass. This criterion is used for substances that may decompose during
drying process, and their mass does not reach constant value.
The second criterion is loss in mass (less than 1 mg) in defined amount of
time.
Automatic finish mode (Auto switch off mode)
MA 3Y operator can choose between various types of finish modes:
-
Automatic 1 (change 1mg/10s)
-
Automatic 2 (change 1mg/25s)
-
Automatic 3 (change 1mg/60s)
-
Automatic 4 (change 1mg/90s)
-
Automatic 5 (change 1mg/120s)
-
Time defined (maximum drying time 99 hours 59 minutes)
-
Manual (on pressing a function key)
-
User defined 1 (set mass change ratio ∆m 0,1-9,9mg and time interval ∆t
max 2,55 sec.
-
User defined 2 (set moisture content change ratio ∆%M over time 60s)
-
Test (enables adjusting finish mode parameters to a tested sample)
Automatic finish mode – user defined criterion (mass/time)
MA 3Y operator determines the limit value for loss of mass and time in which
the loss of mass must not exceed a pre-set value. As the moisture analyzer
reaches the set criterion it automatically finishes the measurement (drying
process).