Accessible responses:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
Mass frame format which is moisture analyzer’s response:
1 2
3
4
5
6
7-15
16 17 18 19 20 21
S I space stability
marker space character mass space
unit
CR LF
Switch off continuous transmission in basic measuring unit
Format: C0 CR LF
Accessible responses:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and in progress
Switch on continuous transmission in current measuring unit
Format: CU1 CR LF
Accessible responses:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Mass frame format which is moisture analyzer’s response:
1 2 3
4
5
6
7-15
16 17 18 19 20 21
S U I stability
marker space character mass space
unit
CR LF
Switch off continuous transmission in current weighing unit
Format: CU0 CR LF
Accessible responses:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and in progress
Set low checkweighing threshold LO
Format: DH_XXXXX CR LF, where: _ - space, XXXXX – mass format
Accessible responses:
DH_OK CR LF - command carried out