Transmille 1000-Series User Manual
Page 89
1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 89
To enable thermocouple simulation to be set up, the following command sequence
should be used:
6) AUTOMATIC COLD JUNCTION COMPENSATION
7) TYPE R
8) 250°C
9) Output ON
Send the following command sequence:
R60/K1/L4/O250/S0
R60 = Thermocouple simulation mode activation
K1 = Automatic cold junction compensation (as detailed in the table above)
L4 = Type R thermocouple
O250 = 250°C output
S0 = Standby OFF (i.e. output switched ON)