1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 89


To enable thermocouple simulation to be set up, the following command sequence
should be used:

////


For example, to set the following configuration:

6) AUTOMATIC COLD JUNCTION COMPENSATION
7) TYPE R
8) 250°C
9) Output ON


Send the following command sequence:

R60/K1/L4/O250/S0

R60 = Thermocouple simulation mode activation
K1 = Automatic cold junction compensation (as detailed in the table above)
L4 = Type R thermocouple
O250 = 250°C output
S0 = Standby OFF (i.e. output switched ON)
= Carriage Return (ASCII character 13)