1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 35
Adjusting the output using the digital control
After the output has been set, any digit of the output display can be incremented or
decremented using either the digital control or the up and down arrow keys.
This function makes calibration of analogue meters easy, where deviating the output
of the calibrator rather than interpreting the indicated value from the UUT provides
more accurate results.
For further information on adjusting the output using the digital control, please refer
to Page 22
Display of % or ppm Error
When the output value is changed by the methods above, the display will show the
change in ppm or % from the original reference value entered from the keyboard.
This feature is ideal for displaying the error in a meter under test by adjusting the
output from the calibrator to make the meter read the nominal value.
For further information on displaying the deviation from nominal, please refer to
Page 22